Multi-parameter Patient Monitor for Infant NIC-YD801
Other Infant Products

Multi-parameter Patient Monitor for Infant NIC-YD801

模型: NIC-YD801

Standard Configuration: 12.1-inch LED, 3/5 Lead ECG, NIBP, SpO2, Pulse Rate, Temperature, Respiration Optional Configuration: 2-Temperature, 1/2 IBP, Digital SpO2, Perfusion Index (PI), Laser Printer Interface Optional Accessories: Touch Screen, Printer, Central Monitoring Station, Neo/Ped Accessories, Wall-mounting, Trolley

产品描述

Features

 

Core

● Newly advanced A9 main board with Linux OS

● Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups NIBP review and 2000 alarm events review

 

Body

● 12.1 inches high-brightness TFT LED

● Support display 9~13 waveforms

● Support 7 channel ECG waveform display simultaneously

● Optional HDMI output

 

Printer

● Built-in high-speed 50mm thermal printer

 

Central System

● Wired or wireless connection

 

Alarm

● Three-level acousto-optic alarm

● Sensor-off alarm

● Paper out alarm

● Support alarm review

● Support alarm pause

 

Linux OS

● Support operation with USB mouse and computer keyboard (option)

● Multi-display mode

● NIBP self-test mode

● Support medical history search by patient ID, name and mobile number

● 13 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection

● Drug calculation and titration table

● Multi-language display

● Online software upgrading by net/USB

 

Performance Specifications

 

Dimension and Weight

● Dimension: 310mm*290mm*150mm

● Weight: 3.4 千克 (excluding accessories)

 

Power Supply

● Voltage: AC100~240V, 50/60HZ, Power≤60W

 

Display

● 12.1" color TFT LED resolution: 800*600 pixels

 

Battery

● Type: Rechargeable lithium battery 12V/2200mAh

● Charge Cycle: ≥500 times

● Working time: 2 hours

 

Recorder (Option)

● Method: Thermal printer

● Paper width: 50 mm (1.97 in)

● Printing speed: 12.5/ 25/ 50 mm/s

● Trace: Max. 3 tracks

● Recording way: Real-time Recording, Review Printing, Periodic Recording, Alarm Recording

 

Alarm

● Level: Low, medium and high

● Indication: Auditory and visual

● Alarm volume adjustable

● Alarm pause time: 1min, 2min

● Parameter alarm type: Latch/ Unlatch

 

Input Device

● Standard: Knob /Keypress

● Option: USB Mouse / USB Keyboard / Touch screen

 

System Output & Extensible Interface

● Ethernet Network: standard RJ45 socket * 1pc

● USB Port: 1pc

● Video Output: HDMI port (option) * 1pc

 

Operating Environment

● Temperature: 5 ~ 40  ℃

● Humidity: 15% ~ 90% (non-condensing)

● Atmosphere pressure: 86 KPa ~ 110 KPa

 

Transportation and Storage

● Temperature:   -20~50  ℃

● Humidity:    10%~90%   (non-condensing)

● Atmosphere pressure:   86 KPa ~ 110 KPa

 

Safety

● IEC60601- 1 Approved, CE marking according to MDD93/42/EEC

● With reference to RoHS Directive 2011/65/EU recasting

 

Trend & Reviewing

● Trend: 2160 hours

● ARR events: 128 groups of ARR events and associated waveform

● NIBP measurement reviewing: 2000 groups

● Waveform review: 2 hours

● Alarm event: 2000 groups of parameter alarms events and associated parameter

 

SpO2

● Measurement Range: 0 ~ 100 %

● Resolution: 1 %

● Accuracy:   ±2% (70% ~ 100%)

                       ±3 % (40% ~ 69%)

                     Unspecified (0 ~ 39%)

● Support Pitch tone and multi-level volume

● User-selectable waveform speed:6.25, 12.5, 25, 50 mm/s

● PI range (Option):0.075%-20%

 

Pulse Rate

● Measuring and Alarm Range: 20~250bpm

● Accuracy: ±1% or ±1 bpm, whichever is greater

● Resolution: 1bpm

 

Respiration

● Method: Impedance between RA-LL, RA-LA

● Gain: ×0.25, ×0.50, ×1, ×2, ×4

● Respiration Rate: 0bpm, 6 ~ 150 BrPM

● Sweep speed:   6.25 mm/s, 12.5 mm/s, 25mm/s

● Resolution: 1 BrPM

Accuracy:±2BrPM or ±2% whichever is greater (7~ 150BrPM)

                 Unspecified (0%~ 6BrPM)

● Apnea Alarm: 10 ~ 40 s

 

Temperature

● Technique: Thermistor probe (2.25K)

● Channel: Dual-channel, provide T1; T2; ΔT

● Measuring and Alarm Range: 0.0  ℃~ 50  ℃ (32℉~ 122℉)

● Unit: Celsius (℃), Fahrenheit (  ℉)

● Resolution: 0.1℃or 1℉

● Accuracy: ±0.1℃(exclusive probe)

 

ECG

● Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V

● Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof

● Gain: 2.5mm/mV(×0.25), 5.0mm/mV(×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto

● Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s

● ECG signal range:   ±5 mV p-p

● Accuracy: ±1bpm/ ±1%, whichever is greater

● Resolution: 1 bpm

● Leakage Current < 10 μA

● Baseline Recovery:≤ 3s after defibrillation (Monitor mode) 

                                      ≤ 1s after defibrillation (Surgery mode)

● Bandwidth:  Surgery 1 ~ 20 Hz           Monitor 0.5 ~ 40 Hz        Diagnostic 0.05 ~ 130 Hz 

● Indication of Electrode Separation: Every electrode (exclusive of RL)

 

Heart Rate

● Measure range: Adult:15~300bpm  Neo/Ped15~350bpm

● Resolution: 1 bpm

● Accuracy: ±1%

 

ST Measurement

● Range: -2.0 ~ +2.0 mV

● Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is - greater

● Other range: unspecified

● Resolution: 0.01mV

 

NIBP

● Method: Oscillometric                                                         -

● Measure mode: Manual, Auto, STAT

● Measure Interval in AUTO Mode 1~480 min

● STAT mode cycle time: Keep 5 minutes, at 5 seconds interval

● Measure and Alarm Range:

                       Adult:  SYS:       40 ~ 280 mmHg

                                   DIA:       10 ~ 220 mmHg

                                   MEAN:   20 ~ 240 mmHg

                 Pediatric:   SYS:       40 ~ 220 mmHg

                                   DIA:       10 ~ 160 mmHg
                                   MEAN:   20 ~ 170 mmHg

                 Neonate:   SYS:        30 ~ 135 mmHg

                                   DIA:       10 ~ 110 mmHg

                                   MEAN:   20 ~ 1 10 mmHg

● Static pressure accuracy: ±3mmHg

● Resolution: 1mmHg

● Accuracy: Maximum Mean error ±5mmHg            

                   Maximum Standard deviation ≤8mmHg

● Over pressure Protection: Dual protection via software & hardware

 

EtCO2 (Mainstream/ Sidestream) (Option)
● Measure method: Non-dispersive infrared (NDIR)
● Measure Range: 0 ~19.7% (0 ~ 150 mmHg)
                             0 ~ 20 kPa
● Resolution: 0.1 mmHg
● CO2 Accuracy:   0 ~ 40 mmHg, ±2 mmHg
                              41 ~ 70 mmHg, ±5% of reading
                              71 ~ 100 mmHg, ±8% of reading
                              101~ 150 mmHg, ±10% of reading
                              at 760 mmHg, ambient temperature of 25℃)
● Respiratory Rate: Range: 3 ~150 BrPM
                               Accuracy: ±1 bpm
 
EtCO2 (Micro-stream) (Option)
● Measure method: Non-dispersive infrared (NDIR)
● Measure Range: 0 ~19.7% (0 ~150 mmHg)
                             0 ~ 20 kPa
● Sample Rate: 50 mL/min ±10mL/min
● Resolution: 0.1 mmHg (0 ~ 50 mmHg)
                      0.25 mmHg (50 ~ 114 mmHg)
● CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg
                             41 ~ 70 mmHg, ±5% of reading
                             71 ~ 100 mmHg, ±8% of reading
                             101~ 150 mmHg, ±10% of reading
                              at 760 mmHg, ambient temperature of 35℃)
● Respiratory Rate: Range: 3 ~120 BrPM
                               Accuracy: ±1 bpm
 
IBP (Option)
● Max Channel: 2
● Measurement way: Thermal resistance way
● Press Sensor: Sensitivity 5 uV/V/mmHg, ±2%
                         Impedance 300 to 3000Ω
● Resolution: 1 mmHg
● Unit: mmKg, kPa, cmH2O
● Transducer sites:   Arterial Pressure (ART)
                                Pulmonary Arterial (PA)
                                Left Arterial (LAP)
                                Right Arterial (RAP)
                                Central Venous Pressure (CVP)
                                Intracranial Pressure (ICP)
                                P1/ P2
● Measuring and alarm range:   ART   0 ~ +350mmHg
                                                  PA       -10 ~ +120 mmHg
                    CVP/ RAP/ LAP/ ICP     -10 ~ +40 mmHg
                                             P1/ P2     -50 ~ +350mmHg
● Accuracy: Static: ±1mmHg or ±2%, whichever is greater (exclusive of transducer)
                               ±4mmHg or ±4%, whichever is greater (inclusive of transducer)
              Dynamic: ±4mmHg or ±4%, whichever is greater