Multi-parameter Patient Monitor for Infant NIC-YD801
模型: NIC-YD801
Standard Configuration: 12.1-inch LED, 3/5 Lead ECG, NIBP, SpO2, Pulse Rate, Temperature, Respiration Optional Configuration: 2-Temperature, 1/2 IBP, Digital SpO2, Perfusion Index (PI), Laser Printer Interface Optional Accessories: Touch Screen, Printer, Central Monitoring Station, Neo/Ped Accessories, Wall-mounting, Trolley
产品描述
Features
Core
● Newly advanced A9 main board with Linux OS
● Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups NIBP review and 2000 alarm events review
Body
● 12.1 inches high-brightness TFT LED
● Support display 9~13 waveforms
● Support 7 channel ECG waveform display simultaneously
● Optional HDMI output
Printer
● Built-in high-speed 50mm thermal printer
Central System
● Wired or wireless connection
Alarm
● Three-level acousto-optic alarm
● Sensor-off alarm
● Paper out alarm
● Support alarm review
● Support alarm pause
Linux OS
● Support operation with USB mouse and computer keyboard (option)
● Multi-display mode
● NIBP self-test mode
● Support medical history search by patient ID, name and mobile number
● 13 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
● Drug calculation and titration table
● Multi-language display
● Online software upgrading by net/USB
Performance Specifications
Dimension and Weight
● Dimension: 310mm*290mm*150mm
● Weight: 3.4 千克 (excluding accessories)
Power Supply
● Voltage: AC100~240V, 50/60HZ, Power≤60W
Display
● 12.1" color TFT LED resolution: 800*600 pixels
Battery
● Type: Rechargeable lithium battery 12V/2200mAh
● Charge Cycle: ≥500 times
● Working time: 2 hours
Recorder (Option)
● Method: Thermal printer
● Paper width: 50 mm (1.97 in)
● Printing speed: 12.5/ 25/ 50 mm/s
● Trace: Max. 3 tracks
● Recording way: Real-time Recording, Review Printing, Periodic Recording, Alarm Recording
Alarm
● Level: Low, medium and high
● Indication: Auditory and visual
● Alarm volume adjustable
● Alarm pause time: 1min, 2min
● Parameter alarm type: Latch/ Unlatch
Input Device
● Standard: Knob /Keypress
● Option: USB Mouse / USB Keyboard / Touch screen
System Output & Extensible Interface
● Ethernet Network: standard RJ45 socket * 1pc
● USB Port: 1pc
● Video Output: HDMI port (option) * 1pc
Operating Environment
● Temperature: 5 ~ 40 ℃
● Humidity: 15% ~ 90% (non-condensing)
● Atmosphere pressure: 86 KPa ~ 110 KPa
Transportation and Storage
● Temperature: -20~50 ℃
● Humidity: 10%~90% (non-condensing)
● Atmosphere pressure: 86 KPa ~ 110 KPa
Safety
● IEC60601- 1 Approved, CE marking according to MDD93/42/EEC
● With reference to RoHS Directive 2011/65/EU recasting
Trend & Reviewing
● Trend: 2160 hours
● ARR events: 128 groups of ARR events and associated waveform
● NIBP measurement reviewing: 2000 groups
● Waveform review: 2 hours
● Alarm event: 2000 groups of parameter alarms events and associated parameter
SpO2
● Measurement Range: 0 ~ 100 %
● Resolution: 1 %
● Accuracy: ±2% (70% ~ 100%)
±3 % (40% ~ 69%)
Unspecified (0 ~ 39%)
● Support Pitch tone and multi-level volume
● User-selectable waveform speed:6.25, 12.5, 25, 50 mm/s
● PI range (Option):0.075%-20%
Pulse Rate
● Measuring and Alarm Range: 20~250bpm
● Accuracy: ±1% or ±1 bpm, whichever is greater
● Resolution: 1bpm
Respiration
● Method: Impedance between RA-LL, RA-LA
● Gain: ×0.25, ×0.50, ×1, ×2, ×4
● Respiration Rate: 0bpm, 6 ~ 150 BrPM
● Sweep speed: 6.25 mm/s, 12.5 mm/s, 25mm/s
● Resolution: 1 BrPM
Accuracy:±2BrPM or ±2% whichever is greater (7~ 150BrPM)
Unspecified (0%~ 6BrPM)
● Apnea Alarm: 10 ~ 40 s
Temperature
● Technique: Thermistor probe (2.25K)
● Channel: Dual-channel, provide T1; T2; ΔT
● Measuring and Alarm Range: 0.0 ℃~ 50 ℃ (32℉~ 122℉)
● Unit: Celsius (℃), Fahrenheit ( ℉)
● Resolution: 0.1℃or 1℉
● Accuracy: ±0.1℃(exclusive probe)
ECG
● Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
● Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof
● Gain: 2.5mm/mV(×0.25), 5.0mm/mV(×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto
● Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
● ECG signal range: ±5 mV p-p
● Accuracy: ±1bpm/ ±1%, whichever is greater
● Resolution: 1 bpm
● Leakage Current < 10 μA
● Baseline Recovery:≤ 3s after defibrillation (Monitor mode)
≤ 1s after defibrillation (Surgery mode)
● Bandwidth: Surgery 1 ~ 20 Hz Monitor 0.5 ~ 40 Hz Diagnostic 0.05 ~ 130 Hz
● Indication of Electrode Separation: Every electrode (exclusive of RL)
Heart Rate
● Measure range: Adult:15~300bpm Neo/Ped15~350bpm
● Resolution: 1 bpm
● Accuracy: ±1%
ST Measurement
● Range: -2.0 ~ +2.0 mV
● Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is - greater
● Other range: unspecified
● Resolution: 0.01mV
NIBP
● Method: Oscillometric -
● Measure mode: Manual, Auto, STAT
● Measure Interval in AUTO Mode 1~480 min
● STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
● Measure and Alarm Range:
Adult: SYS: 40 ~ 280 mmHg
DIA: 10 ~ 220 mmHg
MEAN: 20 ~ 240 mmHg
Pediatric: SYS: 40 ~ 220 mmHg
DIA: 10 ~ 160 mmHg
MEAN: 20 ~ 170 mmHg
Neonate: SYS: 30 ~ 135 mmHg
DIA: 10 ~ 110 mmHg
MEAN: 20 ~ 1 10 mmHg
● Static pressure accuracy: ±3mmHg
● Resolution: 1mmHg
● Accuracy: Maximum Mean error ±5mmHg
Maximum Standard deviation ≤8mmHg
● Over pressure Protection: Dual protection via software & hardware